Products

CA3000N Direct Reading Spectroscopy Special Argon Purifier
  • CA3000N Direct Reading Spectroscopy Special Argon Purifier

CA3000N Direct Reading Spectroscopy Special Argon Purifier

Description:

CA3000NArgon gas purifier is suitable for the production of many spectrometer companies such as Thermoelectric, Spike, GNR, Oxford, OBLF, Bruker, Shimadzu, Chuangxiang, Jiebo, Nake, Yingan, Rayleigh, Jinyibo, Tianrui, Dongfang, etc.Direct reading spectrometer, hydrogen determination instrument, etc. are used in conjunction to enhance the reliability and stability of analytical data.CA3000NIt is especially suitable for use with direct reading spectrometers with nitrogen channels.

Purification Principle

The purifier uses a catalyst to make trace oxygen in argon react with carbon to generate carbon dioxide, thereby removing the impurity oxygen in the argon, and then uses molecular sieves to adsorb and remove impurities such as H2O and CO2 in the argon.Then, the alloy adsorbent is used to remove the remaining inert impurities such as N2 in the argon gas to obtain high-purity argon gas.High-purity argon can be obtained for use by gas-using equipment.

Technical parameters of argon purifier

Requirements for raw material argon

Generally, it is bottled pure argon, liquid argon, and well-controlled pipeline argon.

Impurity requirements: O2≤500PPm, H2O≤500PPm

Output pure gas

Purity: 99.9999%

Impurity content:

O2≤0.2PPm,

H2O≤0.2PPm (i.e. dew point≤85℃),

CO+CO2≤0.1PPm,

The oxides of S and P are ≤0.1PPm.

N2≤0.2PPm

The number of dust particles (≥0.3μm) is 3-5 per liter.

Gas processing volume: 4M3/h.

Equipment size: 360mm×560mm×1000mm

Equipment Features

1) The catalyst has high activity and low operating temperature. High-purity argon can be obtained by purifying bottled argon through this machine, which reduces costs for users.

2) Dual structure, one set is working while the other set is regenerating and standby, so it can provide continuous gas supply for a long time.

3) The regeneration process of the machines pure gas purge is adopted. No hydrogen is needed for regeneration, so no dedicated machine room is required, and the safety is good.

4) CA3000N adopts touch screen, and pressure, temperature, etc. are detected online. The working and regeneration cycles of this system can be adjusted according to the actual usage of the user, which is conducive to saving gas source.

5) CA3000N has an automatic switching function for working and regeneration, which does not require manual operation. The performance is more stable and reliable, avoiding contamination of the system caused by users misoperation.

6) The equipment has built-in alarm systems for over-temperature, over-pressure, cylinder under-pressure, valve failure, etc. to ensure the safety and reliability of the equipment.

7) The device adds a manual switching function.When the gas purity of the system in use is insufficient, but the switching time of the working tower has not yet arrived, you can click the manual switch button to switch to the standby tower, and the system can operate normally again.

8) Added nitrogen removal function to ensure stable data when analyzing nitrogen with direct reading spectrometer with nitrogen channel

9) Added system heating protection device to avoid economic losses to customers caused by damage to thyristors, PLCs and other accessories

10) When the system is working normally, if the mains power is suddenly cut off or the operator fails to shut down the machine normally, the equipment will automatically start the backup power supply and close all programs to ensure normal use of the equipment next time,Truly realize full automatic intelligence